Description
Core Product Summary
The National Instruments PXI-1033 is a five-slot, 3U PXI chassis with an integrated MXI-Express remote controller. It provides a transparent remote link between a desktop PCI Express host or ExpressCard laptop host and the PXI backplane, making it intended for PC-controlled modular test and measurement systems.
The chassis accepts 3U PXI and CompactPCI peripheral modules and provides five peripheral slots. It is AC powered, supplies a 10 MHz reference clock to each peripheral slot, and provides 25 W cooling capacity per slot. Its compact enclosure, AUTO/HIGH fan-speed selector, and optional rack-mount and handle/feet accessories support benchtop, portable, and rack-mounted automated test arrangements.
Key Technical Specifications
| Attribute | Specification |
|---|---|
| Manufacturer | National Instruments (NI) |
| Model | PXI-1033 |
| Product type | PXI chassis with integrated MXI-Express remote controller |
| Backplane format | 3U PXI; accepts 3U PXI and CompactPCI modules |
| Peripheral slots | 5 |
| PXI Express / hybrid slots | 0 |
| Remote-control interface | Integrated MXI-Express controller |
| Host connection options | Desktop PCI Express board or laptop ExpressCard host, depending on chassis kit |
| Maximum backplane system bandwidth | 132 MB/s |
| Sustained MXI-Express throughput | Up to 110 MB/s |
| Slot cooling capacity | 25 W per slot |
| System reference clock | 10 MHz, supplied independently to each peripheral slot |
| Clock accuracy | ±25 ppm over the operating temperature range |
| Maximum slot-to-slot clock skew | 250 ps |
| AC input voltage range | 100–240 VAC; operating range 90–264 VAC |
| AC input frequency | 50/60 Hz; operating range 47–63 Hz |
| AC input current rating | 4–2 A |
| Operating temperature | 0 to 50 °C |
| Relative humidity, operating | 20 to 80%, noncondensing |
| Dimensions | 177 × 257.1 × 212.8 mm (6.97 × 10.12 × 8.38 in.) |
| Weight | 5 kg (11.0 lb) |
| PXI hardware compliance | PXI Hardware Specification Revision 2.2 |
| Mechanical compliance | IEEE 1101.10 |
Engineering Selection Highlights
- The PXI-1033 provides five peripheral slots, so module count must be planned before selecting it for a test-system configuration.
- It accepts 3U PXI and CompactPCI modules, but it has no PXI Express or hybrid slots; PXIe-only modules are not suitable for this backplane.
- Its integrated MXI-Express controller removes the need for a separate embedded controller slot, leaving all five slots available for peripheral modules.
- The documented 25 W per-slot cooling capacity should be compared with the thermal requirement of every installed module; system-level power budgeting may be required.
- The chassis provides a 10 MHz reference clock to each peripheral slot and a PXI trigger bus, supporting synchronized multi-module measurement and stimulus arrangements.
- Its 100–240 VAC universal input allows use with common single-phase AC supplies without manual voltage-range selection.
- The 0 to 50 °C operating range and indoor-use designation define the environmental boundary for cabinet or bench installation.

PXI-1033

PXI-1033
System Compatibility & Integration Requirements
| Item / Dependency | Verified Compatibility or Requirement | Practical Note |
|---|---|---|
| Module format | 3U PXI and CompactPCI modules | Confirm that each installed module is 3U and electrically compatible with the PXI backplane. |
| PXI Express modules | No PXI Express or hybrid slots | Do not plan PXIe or hybrid-module installations in this chassis. |
| Host control | Desktop PCI Express host card or laptop ExpressCard host | The applicable host interface depends on the specific PXI-1033 kit configuration. |
| Remote link | Integrated MXI-Express controller | A compatible MXI-Express host card and cable are required for the remote PC link. |
| Timing and triggering | 10 MHz reference clock and PXI trigger bus | Use the backplane timing resources for inter-module synchronization and handshaking. |
| Power supply | 100–240 VAC, 50/60 Hz | Connect through the AC power cable, which also serves as the main power disconnect. |
| Cooling | 25 W per slot; forced-air circulation | Verify module heat dissipation against the per-slot cooling capacity. |
| Installation environment | Indoor use; Pollution Degree 2; Measurement Category II | Install in a protected indoor electrical environment. |
Detailed Product Overview
The NI PXI-1033 is a compact five-slot PXI mainframe designed around an integrated MXI-Express remote controller. Rather than occupying a peripheral slot with an embedded controller, the chassis provides the controller internally and connects to a PC through a remote MXI-Express link. This arrangement preserves all five backplane slots for PXI or CompactPCI peripheral modules.
The remote link supports up to 110 MB/s sustained throughput, while the chassis backplane specifies a maximum system bandwidth of 132 MB/s. For test-system design, these figures describe the data path between the host computer and the PXI modules; individual application throughput will also depend on the installed modules, host interface, cable configuration, and software architecture.
The PXI-1033 accepts 3U PXI and CompactPCI modules and complies with PXI Hardware Specification Revision 2.2. It does not provide PXI Express or hybrid slots, so it is intended for conventional PXI and CompactPCI module populations rather than PXIe-based systems. This distinction is central to module selection: a visually similar 3U PXIe module cannot be assumed to fit or operate in this chassis.
Timing resources include a built-in 10 MHz reference clock distributed independently to every peripheral slot, with a maximum slot-to-slot skew of 250 ps and a clock accuracy of ±25 ppm across the operating temperature range. The PXI trigger bus supports triggering and handshaking between modules. These features are useful when combining instruments such as digitizers, arbitrary waveform generators, digital I/O, switches, and source-measure units in one coordinated chassis.
Electrically, the chassis uses a universal AC input of 100–240 VAC at 50/60 Hz, with an operating range of 90–264 VAC and 47–63 Hz. The internal supply provides +3.3 V, +5 V, +12 V, and −12 V rails to the backplane, with short-circuit protection on all outputs and defined over-voltage protection ranges. The 25 W per-slot cooling rating is the principal thermal planning value; modules with higher dissipation require an alternative chassis or a revised module configuration.
The enclosure is specified for indoor use, with an operating temperature range of 0 to 50 °C and noncondensing relative humidity of 20 to 80%. Its AUTO fan setting adjusts fan speed according to intake-air temperature, while the HIGH setting provides increased cooling. At 25 °C ambient in AUTO mode, the published sound-pressure level at the operator position is 37.4 dBA; in HIGH mode it is 51.5 dBA.
Dimensions, Mounting & Connection Notes
The standard PXI-1033 enclosure measures 177 mm high, 257.1 mm wide, and 212.8 mm deep, and weighs 5 kg. Installing the optional feet adds 12.7 mm to the height.
The chassis is designed for benchtop, portable, or rack-mounted use. NI offers a side handle and rubber-feet kit under part number 781482-01 and a rack-mount kit under part number 778948-01. The rack-mount mounting holes are M4 and are symmetrical at the front and rear of the chassis.
Cooling air enters at the bottom of the chassis and exhausts along the rear, right side, and top. Module airflow is directed from P1 to P2, from the bottom of a module to its top. The front-panel power switch controls the internal power supply, while the AC power cable provides the main power disconnect.
Applications & System Role
The PXI-1033 functions as the modular instrument enclosure and remote-control backbone for a PC-based PXI test system. Its integrated MXI-Express controller links the chassis backplane to a desktop or laptop host, allowing PXI modules to be operated as remotely controlled measurement, switching, signal-generation, or data-acquisition resources.
With five peripheral slots, the chassis suits compact automated test equipment (ATE) configurations where a separate controller slot would reduce available module capacity. The PXI trigger bus and independently supplied 10 MHz reference clock make it applicable to multi-module systems that require coordinated timing, triggering, or handshaking.
NI identifies automotive, automated test, and electronics applications as relevant contexts for this chassis. Its portable form factor, AC power input, and optional rack-mount hardware also support laboratory, production-test, and field-service test-system arrangements.
FAQ
What is the NI PXI-1033?
The NI PXI-1033 is a five-slot, 3U PXI chassis with an integrated MXI-Express remote controller for PC-controlled PXI test and measurement systems.
How many modules can the PXI-1033 hold?
It provides five peripheral slots for 3U PXI or CompactPCI modules.
Does the PXI-1033 support PXI Express modules?
No. It has no PXI Express or hybrid slots; it accepts 3U PXI and CompactPCI modules.
How does the PXI-1033 connect to a computer?
It uses its integrated MXI-Express controller to provide a remote link to either a desktop PCI Express host card or a laptop ExpressCard host, depending on the chassis kit.
What is the cooling capacity of each slot?
Each slot provides 25 W of cooling capacity. System power and thermal budgeting may be required when installing multiple modules.
What timing features does the chassis provide?
It provides a 10 MHz reference clock independently to each peripheral slot, a ±25 ppm clock accuracy, a maximum slot-to-slot skew of 250 ps, and a PXI trigger bus.
What power input does the PXI-1033 require?
It requires 100–240 VAC at 50/60 Hz; the designed operating range is 90–264 VAC at 47–63 Hz.
Can the PXI-1033 be rack mounted?
Yes. NI offers an optional rack-mount kit, part number 778948-01, and an optional side handle and rubber-feet kit, part number 781482-01.

WhatsApp: +86 16626708626
Email:
Phone: +86 16626708626