Description
Core Product Summary
The NI SCXI-1000 is a four-slot, AC-powered chassis for National Instruments SCXI signal-conditioning modules. It supplies power to installed modules and manages timing, triggering, and signal routing across the SCXIbus.
The chassis accepts up to four SCXI modules and can be daisy-chained with up to eight SCXI-1000 chassis through one multifunction I/O device or module. It offers multiple AC input configurations, including 100, 120, 220, and 240 VAC options, and is configured for the required country power setting.
Key Technical Specifications
| Parameter | Value |
|---|---|
| Manufacturer | National Instruments (NI) |
| Model | SCXI-1000 |
| Product type | Four-slot AC-powered SCXI chassis |
| Module capacity | Up to 4 SCXI modules |
| Bus | SCXIbus |
| Chassis expansion | Up to 8 SCXI-1000 chassis daisy-chained to one multifunction I/O device or module |
| AC input options | 100, 120, 220, or 240 VAC at 50/60 Hz |
| Maximum input current | 0.6 A at 100 VAC; 0.5 A at 120 VAC; 0.25 A at 220 or 240 VAC |
| Power dissipation | 7 W per slot |
| V+ supply | +18.5 to +25 V, 680 mA maximum load |
| V- supply | -18.5 to -25 V, 680 mA maximum load |
| +5 V supply | 4.75 to 5.25 V, 200 mA maximum load |
| Supply ripple | 1.5 V peak-to-peak for V+ and V-; 50 mV peak-to-peak for +5 V |
| Backplane fuses | 1.5 A on V+ and V- supplies |
| Operating temperature | 0 to 50°C |
| Storage temperature | -20 to 70°C |
| Humidity | 10 to 90% RH, noncondensing |
| Maximum altitude | 2,000 m |
| Dimensions including fan | 180 × 195 × 248 mm (H × W × D) |
| Weight | 3.9 kg |
| Chassis addressing | Revision E: Slot 0 jumpers; Revision F and later: DIP switch; default address 0 |
Engineering Selection Highlights
- The SCXI-1000 provides four SCXI module slots, so total channel count and module mix must fit within that capacity.
- It powers installed SCXI modules and routes timing, trigger, and signals across the SCXIbus, making it the central mechanical and electrical platform for the SCXI subsystem.
- The chassis supports daisy-chaining up to eight SCXI-1000 units to one multifunction I/O device or module, which is relevant for higher-channel-count test systems.
- AC input options include 100, 120, 220, and 240 VAC, but the installed power configuration and corresponding fuse must match the local supply.
- The 680 mA V+ and V- backplane ratings and 200 mA +5 V rating define the electrical budget for the installed SCXI modules.
- Chassis addressing differs by hardware revision: revision E uses Slot 0 jumpers, while revision F and later use a DIP switch.
- The 0 to 50°C operating range and 3.9 kg chassis weight should be considered for benchtop, enclosure, or rack-mounted installations.

SCXI-1000

SCXI-1000
System Compatibility & Integration Requirements
| Item / Dependency | Verified Compatibility or Requirement | Practical Note |
|---|---|---|
| SCXI modules | SCXI-series signal-conditioning modules | Install modules in the four available chassis slots. |
| DAQ interface | Multifunction I/O device or SCXI module | Provides the connection between the SCXI system and the host measurement system. |
| Multi-chassis operation | Up to 8 SCXI-1000 chassis daisy-chained | Requires appropriate chassis addressing and a single multifunction I/O device or module. |
| Chassis address | 0 to 31; default address 0 | Revision E uses Slot 0 jumpers; revision F and later use a DIP switch. |
| AC supply | 100, 120, 220, or 240 VAC at 50/60 Hz | Match the configured input range and fuse to the local supply. |
| Software | NI-DAQmx or Traditional NI-DAQ | Used for SCXI system configuration and data acquisition integration. |
Dimensions, Mounting & Connection Notes
- Dimensions including fan: 180 mm height × 195 mm width × 248 mm depth.
- Weight: 3.9 kg.
- The chassis provides four slots for SCXI modules and includes a rear-mounted fan with replaceable filter for forced-air cooling.
- AC input options are selected for the applicable country configuration; the corresponding fuse must match the selected voltage range.
- The SCXIbus backplane supplies V+, V-, and +5 V power to installed modules.
- V+ and V- supplies are fused at 1.5 A on the SCXI-1000 backplane.
- Chassis addressing is set at Slot 0: revision E uses jumpers, and revision F or later uses a DIP switch.
Detailed Product Overview
The NI SCXI-1000 is a four-slot, AC-powered chassis in National Instruments’ SCXI (Signal Conditioning eXtensions for Instrumentation) platform. It provides the enclosure, backplane, power distribution, cooling, and control functions required by SCXI signal-conditioning modules.
Each installed SCXI module receives power from the chassis backplane. The SCXI-1000 provides V+ and V- analog supplies rated at 680 mA each and a +5 V digital supply rated at 200 mA. The backplane also carries the SCXIbus communication, analog, and trigger functions, allowing modules to share timing and routing resources within the chassis.
The chassis supports up to four SCXI modules. For applications requiring additional channels, multiple SCXI-1000 chassis can be daisy-chained; NI documentation specifies a configuration of up to eight SCXI-1000 chassis connected to one multifunction I/O device or module. Chassis addressing distinguishes units in such systems. Revision E chassis use jumpers in Slot 0, while revision F and later chassis use a DIP switch; the default chassis address is 0.
The SCXI-1000 is powered from AC mains and is available in several country-specific input configurations, including 100, 120, 220, and 240 VAC at 50 or 60 Hz. The installed power configuration must match the local supply, and the associated fuse must be appropriate for the selected voltage range.
The chassis includes forced-air cooling through a rear-mounted fan and replaceable filter. Its published environmental limits are 0 to 50°C operating temperature, -20 to 70°C storage temperature, and 10 to 90% relative humidity, noncondensing. The 180 × 195 × 248 mm housing and 3.9 kg weight should be considered when planning benchtop placement, cabinet integration, or rack mounting.
Applications & System Role
The SCXI-1000 is the mechanical, electrical, and signal-routing platform for an NI SCXI signal-conditioning system. It houses up to four SCXI modules, supplies their operating power, and coordinates timing, triggering, and signal routing through the SCXIbus.
In a measurement system, the chassis sits between field signals and the host data-acquisition device. SCXI modules condition signals such as analog inputs, thermocouple measurements, strain-gauge signals, digital I/O, or other instrumentation signals, while the SCXI-1000 provides the shared backplane, power, and expansion architecture.
FAQ
What is the NI SCXI-1000?
It is a four-slot, AC-powered chassis for National Instruments SCXI signal-conditioning modules.
How many SCXI modules does the SCXI-1000 hold?
It holds up to four SCXI modules.
What functions does the chassis provide?
It supplies power to SCXI modules and handles timing, trigger, and signal routing across the SCXIbus.
Can multiple SCXI-1000 chassis be connected together?
Yes. Up to eight SCXI-1000 chassis can be daisy-chained to one multifunction I/O device or module.
What AC input options does it support?
It supports 100, 120, 220, and 240 VAC configurations at 50 or 60 Hz.
How is the chassis address configured?
Revision E chassis use jumpers in Slot 0, while revision F and later chassis use a DIP switch. The default address is 0.
What are its operating environmental limits?
It operates from 0 to 50°C, stores from -20 to 70°C, and operates at 10 to 90% relative humidity, noncondensing.
What are its dimensions and weight?
The chassis measures 180 × 195 × 248 mm including the fan and weighs 3.9 kg.

WhatsApp: +86 16626708626
Email: