Sale!

NI SCXI-1120D 8-Channel Wideband Isolated Analog Input Module

The NI SCXI-1120D is an 8-channel wideband isolated analog input module for the SCXI signal conditioning platform. It provides independent isolation amplifiers on each channel for conditioning volt, millivolt, and current signals in the presence of high common-mode voltages, with bandwidth support up to approximately 20 kHz.

Categories: , , , , SKU: SCXI-1120D Brand:

Description

Core Product Summary

The NI SCXI-1120D is an 8-channel wideband isolated analog input module for the SCXI signal conditioning platform. It provides independent isolation amplifiers on each channel for conditioning volt, millivolt, and current signals in the presence of high common-mode voltages, with bandwidth support up to approximately 20 kHz.

Each channel offers jumper-configurable gain (overall gains from 0.5 to 1000) and a two-stage lowpass filter selectable for 4.5 kHz or 22.5 kHz. The module delivers 250 Vrms working isolation channel-to-channel or channel-to-earth, supports DC or AC coupling (with appropriate terminal blocks), and installs in any SCXI chassis for use with compatible NI data acquisition devices in multiplexed or parallel output modes.

Key Technical Specifications

Parameter Specification
Product type 8-channel wideband isolated analog input module
Channels 8 differential, independently isolated
Overall gains (jumper-selectable) 0.5, 1, 2.5, 5, 10, 25, 50, 100, 125, 250, 500, 1000 (fixed pre-stage gain of 0.5 + two configurable stages)
Input signal ranges (examples) ±10 V (gain 0.5) to ±10 mV (gain 1000); intermediate ranges scale with gain
Lowpass filter (jumper-selectable, two-stage) 4.5 kHz or 22.5 kHz
Maximum working voltage (signal + common-mode) 250 Vrms
Overvoltage protection 250 Vrms (power on or off)
Isolation 250 Vrms working, channel-to-channel or channel-to-earth; CE double-insulated Category II for 250 Vrms operational isolation
Input coupling DC; AC with compatible terminal blocks (e.g., TBX-1329 or SCXI-1305)
Output range ±5 V
Operating temperature 0 to 50 °C
Dimensions 3.1 × 17.3 × 20.3 cm (1.2 × 6.8 × 8.0 in)
Connectors 50-pin male ribbon-cable rear; 32-pin DIN C male front

Detailed Product Overview

The SCXI-1120D functions as a front-end isolation and conditioning stage within an SCXI-based data acquisition system. It converts differential input signals that may ride on high common-mode voltages into single-ended signals referenced to chassis ground, while providing amplification and filtering before the signals reach a controlling DAQ device.

Each of the eight channels contains an isolation amplifier with a fixed pre-stage gain of 0.5 followed by two user-settable gain stages. The first stage offers gains of 1, 10, 50, or 100; the second stage offers 1, 2, 5, 10, or 20. The resulting overall gains span 0.5 to 1000, allowing the module to handle input ranges from ±10 V down to low millivolt levels. This architecture is particularly suited to applications requiring both isolation and moderate-to-high bandwidth, such as production testing, laboratory measurements, and industrial process monitoring of voltage or current sources.

Two-stage lowpass filtering is provided on every channel. Jumpers select either 4.5 kHz or 22.5 kHz cutoff; both stages must be set identically to achieve the intended bandwidth. The module ships configured for 4.5 kHz. The higher bandwidth setting supports signals requiring response up to roughly 20 kHz, distinguishing the SCXI-1120D from the lower-bandwidth SCXI-1120.

Isolation is rated at 250 Vrms working voltage between channels or from channel to earth ground, with continuous overvoltage protection to the same level. The design meets double-insulation Category II requirements for 250 Vrms operational isolation. Input coupling is DC by default; AC coupling is available when the appropriate terminal block is used.

The module mounts in any standard SCXI chassis slot and connects to a DAQ device via the rear 50-pin connector. Front-panel signals interface through the 32-pin DIN C connector, typically via a compatible terminal block that provides screw terminals and, in some cases, cold-junction compensation or attenuation. Output can operate in multiplexed mode (all channels scanned onto one DAQ input) or parallel mode for higher aggregate throughput when the DAQ device supports it.

Configuration is entirely hardware-based via on-board jumpers for gain, filter bandwidth, and certain digital interface options. No onboard programmable gain or filter control is provided; settings are fixed until the module is removed and jumpers repositioned. Compatible terminal blocks extend functionality—for example, high-voltage attenuation or AC coupling—while remaining within the module’s isolation ratings.

SCXI-1120D

SCXI-1120D

SCXI-1120D

SCXI-1120D

Dimensions, Mounting & Connection Notes

  • Dimensions: 3.1 × 17.3 × 20.3 cm (1.2 × 6.8 × 8.0 in)
  • Mounting: Front-panel thumbscrews secure the module to the top and bottom threaded strips of the SCXI chassis.
  • Front connector: 32-pin DIN C male; accepts compatible terminal blocks for field wiring.
  • Rear connector: 50-pin male ribbon-cable for SCXIbus and DAQ device interconnection.
  • Wire preparation (when using terminal blocks): Strip insulation no more than 7 mm; use wire rated for the applied voltage (minimum 300 V / 60 °C for 250 V measurements).
  • Torque on terminal-block screws: 5 to 7 in-lb.
  • Safety earth ground connection required on the terminal block when measuring voltages ≥30 Vrms / 42.4 Vpeak or 60 VDC.

System Compatibility & Integration Requirements

Item / Dependency Verified Compatibility or Requirement Practical Note
Chassis Any SCXI chassis Install in any available slot; power chassis off before insertion
DAQ devices MIO series, 1200 Series, Lab PC+, Lab-NB, Lab-LC, Lab-PC, PC-LPM-16/PnP Full functionality with MIO/1200/Lab PC+; some older devices cannot scan in multiplexed mode
Output modes Multiplexed or parallel Parallel mode increases scanning rates when supported by the DAQ device
Terminal blocks SCXI-1320, SCXI-1327, SCXI-1328, SCXI-1305, TBX-1316, TBX-1329 and related Provide screw terminals, CJC, attenuation, or AC coupling; select according to signal type and voltage
Gain and filter setting Hardware jumpers on the module Must be set before installation; both filter stages must match
Cable Standard SCXI rear cable assembly to DAQ device One DAQ device typically cables to one module per chassis

FAQ

What is the NI SCXI-1120D?
It is an 8-channel wideband isolated analog input module for the SCXI platform that conditions volt, millivolt, and current signals in the presence of high common-mode voltages while providing selectable gain and lowpass filtering.

What bandwidth does the SCXI-1120D support?
Jumper-selectable two-stage lowpass filters provide 4.5 kHz or 22.5 kHz cutoff frequencies, supporting signal bandwidths up to approximately 20 kHz.

What isolation rating does the module provide?
Each channel is isolated with a 250 Vrms working common-mode voltage rating channel-to-channel or channel-to-earth, and the module is rated for continuous overvoltage protection to 250 Vrms.

How is gain configured on the SCXI-1120D?
Gain is set entirely by hardware jumpers. A fixed pre-stage gain of 0.5 combines with two configurable stages to produce overall gains of 0.5, 1, 2.5, 5, 10, 25, 50, 100, 125, 250, 500, and 1000.

Which terminal blocks work with the SCXI-1120D?
Compatible blocks include the SCXI-1320, SCXI-1327, SCXI-1328, SCXI-1305, TBX-1316, and TBX-1329, among others, providing screw-terminal access, optional cold-junction compensation, attenuation, or AC coupling.

Can the SCXI-1120D operate in parallel mode?
Yes. Parallel output mode is supported and can increase scanning rates when the controlling DAQ device is capable of reading multiple channels simultaneously.

What is the operating temperature range?
The module is specified for operation from 0 to 50 °C.