Sale!

NI SCXI-1313A High-Voltage Attenuator Terminal Block

The NI SCXI-1313A is a shielded front-mount terminal block designed specifically for the SCXI-1125 isolated analog input module. It provides eight channels of precision 100:1 resistive attenuation that can be enabled or bypassed on a per-channel basis through software, extending the measurable input range to 150 Vrms or ±150 VDC.

Categories: , , , , SKU: SCXI-1313A Brand:

Description

Core Product Summary

The NI SCXI-1313A is a shielded front-mount terminal block designed specifically for the SCXI-1125 isolated analog input module. It provides eight channels of precision 100:1 resistive attenuation that can be enabled or bypassed on a per-channel basis through software, extending the measurable input range to 150 Vrms or ±150 VDC.

The terminal block includes 18 screw terminals for signal and ground connections and an onboard thermistor cold-junction sensor for thermocouple measurements. It mounts directly to the front of the SCXI-1125 and is rated for CAT II measurement category.

Key Technical Specifications

Parameter Specification
Product type High-voltage attenuator terminal block
Compatible module SCXI-1125
Input channels 8
Attenuation 100:1 or 1:1 (programmable per channel)
Maximum input range 150 Vrms or ±150 VDC
Measurement category CAT II
Cold-junction sensor Thermistor
Sensor accuracy ±0.5 °C (15 to 35 °C); ±0.9 °C (0 to 15 °C and 35 to 55 °C)
Sensor output 1.91 to 0.65 V from 0 to 50 °C
Coupling DC only
Common-mode isolation 150 Vrms or ±150 VDC (channel-to-channel and channel-to-ground)
Screw terminals 18 (8 signal pairs + 2 functional ground)
Maximum wire gauge 16 AWG

Detailed Product Overview

The SCXI-1313A expands the voltage measurement capability of the SCXI-1125 by placing a precision resistive divider in series with each of the eight input channels. When the 100:1 attenuation path is selected, the module can safely measure signals up to 150 Vrms or ±150 VDC while remaining within the input range of the SCXI-1125. For low-voltage applications the dividers can be bypassed individually through software configuration, restoring a 1:1 signal path.

The terminal block is fully shielded and attaches directly to the front connector of the SCXI-1125. Eighteen screw terminals provide convenient field wiring: eight differential pairs map to the module’s analog inputs, and two additional terminals connect to functional chassis ground. A strain-relief bar at the cable entrance helps secure the field wiring.

An onboard thermistor cold-junction sensor supports thermocouple measurements. The sensor produces a voltage output of 1.91 V to 0.65 V over the 0 to 50 °C range and is specified for ±0.5 °C accuracy between 15 and 35 °C. The sensor signal is available to the host module for cold-junction compensation.

Because attenuation is controlled programmatically, the user simply configures the desired input limits in software; the driver automatically selects the correct attenuation path and applies the corresponding scaling. This approach eliminates the need for manual switches or jumpers when changing between high-voltage and low-voltage ranges.

The SCXI-1313A is intended for indoor use in CAT II environments and accepts field wire up to 16 AWG.

System Compatibility & Integration Requirements

Item / Dependency Verified Compatibility or Requirement Practical Note
Host module only Direct front-mount attachment
Attenuation control Software-selectable (100:1 or 1:1) per channel Configured via input limits in NI-DAQmx or MAX
Cold-junction compensation Onboard thermistor sensor Available for thermocouple measurements
Field wiring Up to 16 AWG Use wire rated for the applied voltage
Measurement category CAT II Suitable for local-level mains-connected circuits
Coupling DC only No AC-coupling option on this terminal block

Dimensions, Mounting & Connection Notes

  • Mounting: Direct front attachment to the SCXI-1125 module.
  • Field connections: 18 screw terminals (8 differential pairs + 2 ground).
  • Maximum wire size: 16 AWG.
  • Strain relief: Integrated bar at the terminal-block entrance.
  • Isolation: 150 Vrms / ±150 VDC channel-to-channel and channel-to-ground.
SCXI-1313A

SCXI-1313A

SCXI-1313A

SCXI-1313A

FAQ

What is the NI SCXI-1313A?
It is a shielded high-voltage attenuator terminal block for the SCXI-1125 that provides programmable 100:1 or 1:1 attenuation on each of eight channels and includes an onboard cold-junction sensor.

What voltage range does the SCXI-1313A support?
With the 100:1 attenuators enabled, the terminal block extends the measurable range to 150 Vrms or ±150 VDC.

How is attenuation selected?
Attenuation is selected programmatically on a per-channel basis by setting the appropriate input limits in software; no hardware switches are required.

Does the SCXI-1313A include cold-junction compensation?
Yes. It contains a thermistor cold-junction sensor with specified accuracy of ±0.5 °C from 15 to 35 °C.

Which module is the SCXI-1313A designed for?
It is designed exclusively for use with the SCXI-1125 isolated analog input module.

What is the measurement category rating?
The terminal block is rated CAT II.