Description
Core Product Summary
The NI SCXI-1327 is a shielded high-voltage attenuator terminal block that mounts to the front of compatible SCXI analog input modules. Each of its eight channels contains a precision 100:1 resistive voltage divider that can be enabled or disabled by onboard switches, extending the measurable input range to 300 Vrms or ±300 VDC.
The terminal block provides 18 screw terminals for convenient field wiring and includes an onboard thermistor cold-junction sensor for thermocouple measurements. It is designed for use with the SCXI-1120, SCXI-1120D, SCXI-1121, SCXI-1125, and SCXI-1126 modules and reduces the effective input impedance of the host module to 1 MΩ when the attenuators are engaged.
Key Technical Specifications
| Parameter | Specification |
|---|---|
| Product type | High-voltage attenuator terminal block |
| Compatible modules | SCXI-1120, SCXI-1120D, SCXI-1121, SCXI-1125, SCXI-1126 |
| Attenuation | 100:1 or 1:1 (switch-selectable per channel) |
| Maximum input range | 300 Vrms or ±300 VDC |
| Measurement category | CAT II |
| Cold-junction sensor | Thermistor |
| Sensor output | 1.91 V (0 °C) to 0.58 V (55 °C) |
| Sensor accuracy (typical) | ±0.9 °C (15–35 °C range reference); higher outside this band |
| Screw terminals | 18 (8 signal pairs + ground) |
| Input impedance (with attenuation) | 1 MΩ |
| Coupling | DC |
Detailed Product Overview
The SCXI-1327 expands the voltage-handling capability of selected SCXI isolated analog input modules by inserting a precision resistive divider network in each signal path. When the 100:1 attenuation path is selected via the internal switches, signals up to 300 Vrms or ±300 VDC can be measured while remaining within the safe input range of the host module. For low-voltage work the dividers can be bypassed on a per-channel basis, restoring a direct 1:1 connection.
Eighteen screw terminals provide straightforward field connections: eight differential pairs map to the module inputs (or to a combination of inputs and excitation channels when used with the SCXI-1121), and additional terminals connect to chassis ground. The terminal block is fully shielded and includes a strain-relief feature at the cable entry.
An onboard thermistor cold-junction sensor supports thermocouple measurements. The sensor output voltage ranges from approximately 1.91 V at 0 °C to 0.58 V at 55 °C and is available to the host module for cold-junction compensation. Because the attenuation setting is hardware-based, the corresponding scaling factor must be configured in software (Measurement & Automation Explorer or the application) so that readings are correctly scaled.
The SCXI-1327 is rated for CAT II environments and is intended for indoor use. When the attenuators are engaged, the input impedance presented to the field signals is 1 MΩ.
Dimensions, Mounting & Connection Notes
- Mounting: Direct front attachment to compatible SCXI modules.
- Field connections: 18 screw terminals.
- Attenuation control: Internal per-channel switches.
- Grounding: Dedicated functional ground terminals provided.
- Environment: Indoor use; operating temperature range consistent with SCXI terminal-block family (typically 0 to 50 °C).

SCXI-1327

SCXI-1327
System Compatibility & Integration Requirements
| Item / Dependency | Verified Compatibility or Requirement | Practical Note |
|---|---|---|
| Host modules | SCXI-1120, SCXI-1120D, SCXI-1121, SCXI-1125, SCXI-1126 | Direct front-mount attachment |
| Attenuation selection | Onboard switches (100:1 or 1:1) per channel | Must be matched by software scaling configuration |
| Cold-junction compensation | Onboard thermistor sensor | Available for thermocouple applications |
| Field wiring | Screw terminals | Use wire rated for the applied voltage |
| Input impedance | 1 MΩ when attenuators are enabled | Affects source loading |
| Measurement category | CAT II, 300 V working voltage | Suitable for local-level circuits |
FAQ
What is the NI SCXI-1327?
It is a shielded high-voltage attenuator terminal block that provides switchable 100:1 attenuation on each of eight channels and an onboard cold-junction sensor for use with selected SCXI analog input modules.
What voltage range does the SCXI-1327 support?
With the 100:1 attenuators enabled, the terminal block extends the measurable range to 300 Vrms or ±300 VDC.
How is attenuation selected?
Attenuation is selected by onboard switches on a per-channel basis. The corresponding scaling must also be set in software.
Which modules are compatible with the SCXI-1327?
It is compatible with the SCXI-1120, SCXI-1120D, SCXI-1121, SCXI-1125, and SCXI-1126.
Does the SCXI-1327 include cold-junction compensation?
Yes. It contains a thermistor cold-junction sensor whose output is available to the host module.
What is the input impedance when attenuation is enabled?
The effective input impedance is 1 MΩ.

WhatsApp: +86 16626708626
Email: