Description
Core Product Summary
The Yokogawa SC41-SP34 is a 2-electrode contacting conductivity sensor intended for low-conductivity aqueous processes. It belongs to Yokogawa’s SC41 specific-conductivity sensor family and provides conductivity measurement with integrated temperature compensation through an Ni100 temperature element.
The SC41-SP34 uses stainless steel AISI 316L wetted parts, a nominal cell constant of 0.01 cm⁻¹, and a plug-in mounting arrangement. Its 164 mm sensor length and compatibility with Yokogawa FF40 flow fittings and FD40 immersion fittings make it suitable for inline monitoring of purified water, demineralized water, reverse-osmosis permeate, and comparable low-ionic-content streams. The sensor connects to compatible Yokogawa conductivity analyzers or transmitters through the applicable WU40 or WU10/WE10 sensor-cable arrangement.
Key Technical Specifications
| Item | Verified Specification |
|---|---|
| Manufacturer | Yokogawa |
| Model | SC41-SP34 |
| Product type | Contacting conductivity sensor |
| Measurement principle | 2-electrode conductivity measurement |
| Sensor family | Yokogawa SC41 specific-conductivity sensor series |
| Temperature element | Ni100 |
| Nominal cell constant | 0.01 cm⁻¹ |
| Wetted body material | Stainless steel AISI 316L |
| Mounting style | Plug-in |
| Sensor length | 164 mm |
| Compatible flow fitting | Yokogawa FF40 flow fitting |
| Compatible immersion fitting | Yokogawa FD40 immersion fitting |
| Cable options | WU40 cable, or WU10/WE10 cable |
Engineering Selection Highlights
- The SC41-SP34 uses a 0.01 cm⁻¹ nominal cell constant, which is appropriate for low-conductivity water measurement where a small cell constant supports measurement at the low end of the conductivity range.
- Its 2-electrode design places it in Yokogawa’s SC41/SC42-SP contacting-sensor group, so it should be selected for relatively clean aqueous streams rather than heavily fouling or high-conductivity services.
- The integrated Ni100 temperature element provides measurement compensation; it is not positioned by Yokogawa as a process-temperature-control element.
- Stainless steel AISI 316L wetted parts support use in water-based applications where that material is chemically appropriate.
- The plug-in, 164 mm-long construction is designed for installation in a matching Yokogawa FF40 flow fitting or FD40 immersion fitting.
- The SC41 family uses an Ni100 element, whereas the related SC42 family uses Pt1000; analyzer compatibility with the temperature-element type is therefore a key replacement check.
System Compatibility & Integration Requirements
| Item / Dependency | Verified Compatibility or Requirement | Practical Note |
|---|---|---|
| Analyzer / transmitter | Yokogawa contacting-conductivity analyzer or transmitter | Confirm that the installed instrument accepts an SC41 sensor with an Ni100 temperature element. |
| Temperature element | Ni100 | The SC41 differs from the SC42, which uses a Pt1000 temperature element. |
| Flow installation | Yokogawa FF40 flow fitting | Use an FF40 fitting for inline flow-through installation. |
| Immersion installation | Yokogawa FD40 immersion fitting | Use an FD40 fitting for tank, open-vessel, or drain installation. |
| Sensor cable | WU40 cable, or WU10/WE10 cable | Select the cable arrangement specified for the connected Yokogawa analyzer and installation. |
| Extended cable run | WF10 extension cable with BA10 connection box | This arrangement supports cable runs up to 60 m. |
Dimensions, Mounting & Connection Notes
| Item | Verified Detail |
|---|---|
| Sensor length | 164 mm |
| Mounting method | Plug-in installation in a matching Yokogawa fitting |
| Flow fitting | FF40 flow fitting |
| Immersion fitting | FD40 immersion fitting |
| Cable connection | WU40 cable or WU10/WE10 cable |
| Extended cable arrangement | WF10 extension cable with BA10 connection box, up to 60 m |
For immersion installations in agitated tanks or fast-flowing processes, Yokogawa documentation directs that the fitting be adequately supported.

SC41-SP34

SC41-SP34
Detailed Product Overview
The Yokogawa SC41-SP34 is a plug-in, 2-electrode contacting conductivity sensor designed for specific-conductivity measurement in low-conductivity aqueous processes. It uses stainless steel AISI 316L for its wetted body and electrodes and incorporates an Ni100 temperature element for measurement compensation.
The -SP34 suffix identifies a sensor configuration with a nominal cell constant of 0.01 cm⁻¹ and a 164 mm length. In a contacting conductivity cell, the measured solution forms the electrical path between electrodes; the cell constant relates the electrode geometry to the conductivity calculation. A 0.01 cm⁻¹ cell constant is suited to low-ionic-content water, where higher cell constants are generally directed toward more conductive solutions.
The SC41-SP34 is intended to be installed in a matching Yokogawa fitting rather than directly threaded into process piping. For continuous inline measurement, the FF40 flow fitting provides a flow-through installation. For tanks, open vessels, or drains, the FD40 immersion fitting provides the corresponding installation method. Where the fitting is exposed to agitation or rapid flow, adequate mechanical support of the fitting is required.
Electrical integration is through a WU40 cable or a WU10/WE10 cable arrangement, depending on the connected Yokogawa analyzer or transmitter configuration. Where a longer cable run is required, the documented arrangement uses a WF10 extension cable with a BA10 connection box for runs up to 60 m. Because the SC41 uses an Ni100 temperature element, replacement planning should confirm that the existing Yokogawa conductivity instrument supports that element type; the related SC42 sensor family uses Pt1000 instead.
Applications & System Role
The Yokogawa SC41-SP34 sits at the field-measurement level of a liquid-analysis loop. It senses solution conductivity and supplies the measurement signal, including temperature compensation, to a compatible Yokogawa conductivity analyzer or transmitter.
Its 0.01 cm⁻¹ cell constant and contacting 2-electrode construction make it a fit scenario for low-conductivity water streams such as purified water, demineralized water, reverse-osmosis permeate, and feedwater monitoring. In documented Yokogawa low-conductivity-water testing, SC41-SP34 sensors were installed in FF40 flow fittings to monitor incoming and outgoing water conductivity.
FAQ
What is the Yokogawa SC41-SP34?
The Yokogawa SC41-SP34 is a 2-electrode contacting conductivity sensor with an Ni100 temperature element, designed for low-conductivity aqueous applications.
What cell constant does the SC41-SP34 use?
It has a nominal cell constant of 0.01 cm⁻¹.
What materials are used in the SC41-SP34?
The sensor’s wetted body and electrodes are stainless steel AISI 316L.
How is the SC41-SP34 mounted?
It is a plug-in sensor intended for installation in a Yokogawa FF40 flow fitting or FD40 immersion fitting.
What is the sensor length?
The SC41-SP34 sensor length is 164 mm.
Which temperature element does it use?
It uses an Ni100 temperature element for measurement compensation.
How does it differ from the Yokogawa SC42?
The SC41 uses an Ni100 temperature element, while the SC42 uses a Pt1000 temperature element.
Which cables can be used with the SC41-SP34?
It can be connected using a WU40 cable or a WU10/WE10 cable; longer runs up to 60 m can use a WF10 extension cable with a BA10 connection box.

WhatsApp: +86 16626708626
Email:
Phone: +86 16626708626