Description
Core Product Summary
The GE Panametrics XMTC-62-21 is an online thermal conductivity gas transmitter for measuring the concentration of one gas in a binary gas mixture. It uses thermal-conductivity measurement with glass-coated thermistor sensor elements and microprocessor-based signal processing.
The transmitter requires a sample system and provides an isolated, field-programmable 4–20 mA output proportional to the measured gas concentration. It operates from 24 VDC, includes RS232 communication, and is supplied with a 10 ft (3 m) four-wire power and output cable. The XMTC-62-21 is intended for applications involving hydrogen, carbon dioxide, methane, or helium in binary mixtures, with the exact measurement range and calibration defined by the ordered configuration.
Key Technical Specifications
| Category | Specification |
|---|---|
| Product type | Thermal conductivity gas transmitter |
| Product family | GE Panametrics XMTC |
| Measurement principle | Thermal conductivity |
| Measured media | Binary gas mixtures |
| Target gases | Hydrogen, carbon dioxide, methane, or helium, depending on ordered configuration |
| Sensor technology | Glass-coated thermistor elements |
| Analog output | Isolated 4–20 mA, field-programmable |
| Maximum output load | 800 Ω |
| Digital communication | RS232 |
| Supply voltage | 24 VDC ±2 VDC |
| Maximum supply current | 1.2 A at power-up |
| Standard cable | 10 ft (3 m), four-wire color-coded cable |
| Optional cable length | Up to 4,000 ft (1,200 m) |
| Standard sample flow | 0.5 SCFH (250 cc/min) nominal |
| Flow range | 0.01–4.0 SCFH (5–2,000 cc/min) |
| Response time | 20 seconds for 90% of a step change |
| Linearity | ±1% of span |
| Repeatability | ±0.5% of span per week |
| Cell operating temperature | 131°F (55°C) standard; 149°F (65°C) optional |
| Ambient temperature for standard cell | Up to 122°F (50°C) |
| Ambient temperature for optional cell | Up to 149°F (65°C) |
| Weatherproof enclosure dimensions | 9.53 in. high × 5.71 in. diameter (242 × 145 mm) |
| Explosion-proof enclosure dimensions | 10.47 in. high × 5.7 in. diameter (266 × 145 mm) |
| Weight | 9.5 lb (4.3 kg) |
| Electrical connection | 3/4 in. NPTF conduit |
| Sample connections | 1/4 in. NPTF sample inlet/outlet and optional reference inlet/outlet |
Engineering Selection Highlights
- The XMTC-62-21 measures one gas in a binary or pseudobinary mixture; the gas pair and measurement range must be specified in the ordered configuration.
- Its thermal-conductivity sensing approach is applicable to mixtures containing hydrogen, carbon dioxide, methane, or helium.
- A sample system is mandatory: it must deliver a clean, representative sample at a temperature, pressure, and flow rate within the transmitter’s limits.
- The standard sample condition is 0.5 SCFH (250 cc/min) at atmospheric pressure, with the cell operating at 131°F (55°C).
- The isolated 4–20 mA output is field-programmable and supports a maximum load of 800 Ω.
- RS232 communication is provided for transmitter setup and communication with compatible devices.
- The 24 VDC supply must support up to 1.2 A at power-up; GE’s optional PS5R-C24 power supply converts 100/120/220/240 VAC to the required 24 VDC.

XMTC-62-21

XMTC-62-21
System Compatibility & Integration Requirements
| Item / Dependency | Verified Compatibility or Requirement | Practical Note |
|---|---|---|
| Measurement system | GE Panametrics XMTC thermal-conductivity transmitter | The XMTC-62-21 is a self-contained transmitter with integrated sensor and electronics. |
| Sample system | Required | Must provide a clean, representative sample within the specified temperature, pressure, and flow limits. |
| Gas composition | Binary or pseudobinary mixture totaling 100% | The measured gas pair and range are configuration-dependent. |
| Analog control system | Isolated 4–20 mA output | Maximum field load is 800 Ω. |
| Digital communication | RS232 | Used for transmitter communication and configuration. |
| Power supply | 24 VDC ±2 VDC, 1.2 A maximum at power-up | GE’s optional PS5R-C24 power supply is documented for this purpose. |
| Remote display | GE TMO2D-TC display optional | Provides a two-line × 24-character backlit LCD with programming and control capability. |
| Explosion-proof display | GE XDP display optional | Includes a stabilized 24 VDC supply, 3-digit display, two SPDT alarm relays, and isolated 4–20 mA output. |
| Wiring | Standard four-wire color-coded cable | Standard length is 10 ft (3 m); optional lengths extend to 4,000 ft (1,200 m). |
Dimensions, Mounting & Connection Notes
| Item | Verified data |
|---|---|
| Weatherproof enclosure | 9.53 in. high × 5.71 in. diameter (242 × 145 mm) |
| Explosion-proof enclosure | 10.47 in. high × 5.7 in. diameter (266 × 145 mm) |
| Weight | 9.5 lb (4.3 kg) |
| Electrical conduit connection | 3/4 in. NPTF |
| Sample inlet / outlet | 1/4 in. NPTF |
| Optional reference inlet / outlet | 1/4 in. NPTF |
| Standard wiring cable | 10 ft (3 m), four-wire, color-coded |
| Optional cable length | Up to 4,000 ft (1,200 m) |
| Sensor wetted materials | Standard: 316 stainless steel, glass, and Viton O-rings; optional: Hastelloy C276, titanium, and Chemraz O-rings |
Detailed Product Overview
The GE Panametrics XMTC-62-21 is a compact, online thermal conductivity transmitter for binary gas measurement. It combines a thermal-conductivity sensor with associated electronics in a self-contained transmitter package. The measurement principle is based on the thermal conductivity of the sample gas, allowing the transmitter to determine the concentration of one gas in a binary mixture.
The transmitter requires a sample system. The sample system must deliver a clean, representative sample at a temperature, pressure, and flow rate within the XMTC’s acceptable limits. The documented standard sample condition is a flow rate of 0.5 SCFH (250 cc/min) at atmospheric pressure, with a cell operating temperature of 131°F (55°C). Higher-temperature and higher-pressure options are available.
The XMTC-62-21 provides an isolated 4–20 mA analog output proportional to the measured gas concentration. The output is field-programmable and supports a maximum load of 800 Ω. The transmitter also provides RS232 communication for setup and communication with compatible devices. A standard 10 ft (3 m), four-wire, color-coded cable connects the transmitter for power and output signals; optional cables are available in lengths up to 4,000 ft (1,200 m).
The transmitter requires 24 VDC power with a maximum start-up current of 1.2 A. GE offers the PS5R-C24 power supply to convert 100/120/220/240 VAC to the required 24 VDC. Optional display and interface equipment includes the TMO2D-TC remote display and the XDP explosion-proof display.
The XMTC transmitter is available in weatherproof and explosion-proof packaging. The weatherproof enclosure measures 9.53 × 5.71 in. (242 × 145 mm), while the explosion-proof enclosure measures 10.47 × 5.7 in. (266 × 145 mm). The transmitter weighs 9.5 lb (4.3 kg). Electrical connection uses 3/4 in. NPTF conduit, while sample and optional reference connections use 1/4 in. NPTF fittings.
Documented performance includes linearity of ±1% of span and repeatability of ±0.5% of span per week. The response time is 20 seconds for 90% of a step change. The exact measurement range, gas pair, calibration, and hazardous-area configuration depend on the specific XMTC configuration ordered.
Applications & System Role
The XMTC-62-21 is an online gas-analysis transmitter in a process or generator monitoring system. It measures the concentration of a target gas in a binary mixture and converts the thermal-conductivity result into an isolated 4–20 mA signal for a PLC, DCS, recorder, or other control-system input.
The transmitter is relevant to applications involving hydrogen, carbon dioxide, methane, or helium in binary mixtures. A mandatory sample system prepares and delivers the gas sample to the transmitter at the required flow, temperature, and pressure conditions.
Typical fit scenarios are process-gas monitoring, gas purity measurement, and binary gas concentration measurement where a continuous 4–20 mA signal and RS232 communication are required. The exact gas pair, measurement range, calibration, and enclosure configuration are determined by the ordered XMTC configuration.
FAQ
What is the GE XMTC-62-21?
The GE XMTC-62-21 is a Panametrics online thermal conductivity gas transmitter for measuring the concentration of one gas in a binary gas mixture.
Which gases can the XMTC-62-21 measure?
It is applicable to binary mixtures containing hydrogen, carbon dioxide, methane, or helium. The exact gas pair and measurement range depend on the ordered configuration.
What output signal does the XMTC-62-21 provide?
It provides an isolated, field-programmable 4–20 mA output with a maximum load of 800 Ω.
Does the XMTC-62-21 require a sample system?
Yes. A sample system is mandatory and must deliver a clean, representative sample within the transmitter’s temperature, pressure, and flow limits.
What sample flow rate does the XMTC require?
The standard sample flow rate is 0.5 SCFH (250 cc/min) at atmospheric pressure. The documented flow range is 0.01–4.0 SCFH (5–2,000 cc/min).
What power supply does the XMTC-62-21 use?
It requires 24 VDC ±2 VDC and can draw up to 1.2 A at power-up. GE’s optional PS5R-C24 power supply converts 100/120/220/240 VAC to 24 VDC.
What communication interface is included?
The transmitter includes RS232 communication.
What are the transmitter’s performance specifications?
Documented performance includes ±1% of span linearity, ±0.5% of span per week repeatability, and a 20-second response time for 90% of a step change.

WhatsApp: +86 16626708626
Email:
Phone: +86 16626708626