Sale!

NI PXIe-1082 8-Slot 3U PXI Express Chassis

The National Instruments PXIe-1082 is an eight-slot, 3U PXI Express chassis designed for high-performance modular test, measurement, and control systems. It accepts PXI Express modules in every peripheral position and supports hybrid-compatible standard PXI modules in its four hybrid slots.

Categories: , , , , SKU: PXIe-1082 Brand:

Description

Core Product Summary

The National Instruments PXIe-1082 is an eight-slot, 3U PXI Express chassis designed for high-performance modular test, measurement, and control systems. It accepts PXI Express modules in every peripheral position and supports hybrid-compatible standard PXI modules in its four hybrid slots.

The chassis provides one system-controller slot with three expansion positions, four hybrid peripheral slots, two PXI Express peripheral slots, and one PXI Express system timing slot. Its backplane supports up to 8 GB/s system bandwidth and up to 2 GB/s single-direction dedicated bandwidth per PXI Express slot. The AC-powered PXIe-1082 provides 38 W cooling per slot, operates from 0 °C to 55 °C, and includes PXI_CLK10, PXIe_CLK100, PXIe_SYNC100, PXI trigger, star-trigger, and PXIe differential star-trigger resources.

 

Key Technical Specifications

Attribute Specification
Manufacturer National Instruments (NI)
Model PXIe-1082
Product type 3U PXI Express chassis
Part numbers 860717-01; 780321-01
Total slots 8
System controller slot 1, with three controller expansion slots
Hybrid peripheral slots 4, slots 5–8
PXI Express peripheral slots 2, slots 2–3
PXI Express system timing slot 1, slot 4
Standard PXI-only peripheral slots 0
Backplane compliance PXI-5 PXI Express Hardware Specification
Maximum system bandwidth 8 GB/s
PXI Express slot bandwidth Up to 2 GB/s single-direction, x4 PCIe, per slot
Slot cooling capacity 38 W per slot
Maximum total usable power 420 W
Maximum system-slot dissipation 140 W
Maximum peripheral-slot dissipation 38.25 W
Internal PXI_CLK10 stability ±25 ppm
Internal PXIe_CLK100 stability ±25 ppm
PXI_CLK10 slot-to-slot skew Less than 500 ps
PXIe_CLK100 slot-to-slot skew Less than 100 ps
PXI star-trigger slot-to-slot skew 250 ps maximum
PXI Express differential star-trigger skew 150 ps maximum
AC input voltage range 100–240 VAC; operating range 90–264 VAC
AC input frequency 50/60 Hz; operating range 47–63 Hz
AC input current rating 8 A to 4 A
Operating temperature 0 °C to 55 °C
Operating humidity 10% to 90% RH, noncondensing
Storage temperature −40 °C to 71 °C
Dimensions 177.1 × 271.4 × 396.5 mm (6.97 × 10.68 × 15.61 in.)
Weight 8.8 kg (19.4 lb)
Mechanical compliance IEEE 1101.10
External PXI trigger access No additional module required? No
External PXI_CLK10 import/export Yes

 

Engineering Selection Highlights

  • The PXIe-1082 provides eight total slots, but only seven peripheral slots; slot 1 is reserved for the system controller and its expansion positions.
  • Its four hybrid slots accept PXI Express, CompactPCI Express Type-2, hybrid-compatible PXI, and 32-bit CompactPCI peripheral modules.
  • The two PXI Express peripheral slots accept PXI Express and CompactPCI Express Type-2 modules but do not accept standard PXI modules.
  • The system timing slot accepts a PXI Express system timing module or a PXIe peripheral, and provides differential star-trigger connectivity to hybrid slots.
  • The chassis provides 38 W cooling per slot, so each module’s power dissipation must be checked against the slot and total-system power limits.
  • PXI_CLK10, PXIe_CLK100, PXIe_SYNC100, eight PXI trigger lines, and PXIe differential star triggers support synchronized multi-module systems.
  • The rear-panel 10 MHz REF IN/REF OUT BNC connectors and system-timing-slot clock input support external clock distribution and multi-chassis synchronization.
PXIe-1082

PXIe-1082

PXIe-1082

PXIe-1082

System Compatibility & Integration Requirements

Item / Dependency Verified Compatibility or Requirement Practical Note
Module format 3U PXI Express, CompactPCI Express Type-2, hybrid-compatible PXI, and CompactPCI modules Slot type determines which module formats are accepted.
System controller PXI Express system controller in slot 1 The controller can expand left into three dedicated expansion slots without consuming peripheral slots.
Hybrid slots Slots 5–8 These accept PXIe, CompactPCI Express Type-2, hybrid-compatible PXI, and 32-bit CompactPCI modules.
PXI Express slots Slots 2–3 These accept PXIe and CompactPCI Express Type-2 modules only.
System timing slot Slot 4 Accepts a PXI Express system timing module or PXIe peripheral and provides PXIe_DSTAR connectivity.
Standard PXI modules Hybrid-compatible modules only Standard PXI modules must be hybrid-compatible to install in slots 5–8.
Timing resources PXI_CLK10, PXIe_CLK100, PXIe_SYNC100 Each signal is distributed to every peripheral slot through an independent buffer.
Triggering Eight PXI trigger lines, PXI star trigger, PXIe_DSTAR The PXI trigger bus is shared by all slots; differential star triggers connect the timing slot to hybrid slots.
External clocking Rear-panel 10 MHz REF IN and REF OUT; system-timing-slot PXI_CLK10_IN An external 10 MHz clock causes the backplane clocks to phase-lock automatically.
Power 100–240 VAC, 50/60 Hz The AC power cable is the main power disconnect; the rear circuit breaker is push-to-reset.
Cooling Forced-air cooling; 38 W per slot Maintain the specified clearances and install filler panels in unused slots.
Software PXI Platform Services 2.0 or higher Included chassis software media specifies this minimum version.

 

Dimensions, Mounting & Connection Notes

The PXIe-1082 measures 177.1 mm high, 271.4 mm wide, and 396.5 mm deep, and weighs 8.8 kg. It is designed for benchtop or rack-mounted operation and complies with the IEEE 1101.10 mechanical packaging standard.

Cooling air enters through the filtered rear intake and exhausts primarily through the top of the chassis; secondary intake and exhaust vents are located along the sides. NI specifies minimum clearances of 44.45 mm on each side and 76.20 mm at the rear. Unused slots should be fitted with the supplied filler panels, and optional slot blockers can improve thermal performance when slots remain empty.

The rear panel provides a universal AC input, push-to-reset circuit breaker, chassis ground screw, 10 MHz REF OUT and REF IN BNC connectors, remote inhibit and voltage-monitoring connector, fan-speed selector, and inhibit-mode selector. The modular power-supply shuttle is field replaceable.

The front panel includes the power-inhibit switch and LED, removable feet, carry handle, backplane connectors, controller expansion slots, system-controller slot, hybrid slots, system timing slot, and PXI Express peripheral slots. Optional front and rear rack-mount kits are available.

 

Detailed Product Overview

The NI PXIe-1082 is an eight-slot PXI Express mainframe that combines a high-bandwidth backplane, AC power system, forced-air cooling, and PXI timing resources in a 3U chassis. It is designed around the PXI-5 PXI Express Hardware Specification and supports a mixed ecosystem of PXI Express, CompactPCI Express, hybrid-compatible PXI, and CompactPCI modules.

Slot 1 is the system-controller slot. It uses a four-link configuration and includes three expansion positions to the left, allowing wider controllers to occupy expansion space without reducing the number of peripheral slots. The backplane routes three of the system slot’s x4 PCIe links through PCIe switches to the peripheral slots, while slot 2 connects directly to the remaining system-slot link.

Slots 2 and 3 are PXI Express peripheral slots. They accept PXI Express peripheral modules with x4 or x1 PCIe links and CompactPCI Express Type-2 peripheral modules. Slot 4 is the PXI Express system timing slot. In addition to accepting a system timing module or PXIe peripheral, it provides three dedicated differential PXIe_DSTAR pairs to each hybrid peripheral slot and a single-ended PXI star trigger to every slot.

Slots 5 through 8 are hybrid peripheral slots. They provide full PXI Express functionality and 32-bit PXI functionality, except that PXI Local Bus connectivity is limited to Local Bus 6 left and right. These slots accept PXI Express modules, CompactPCI Express Type-2 modules, hybrid-compatible PXI modules using an XJ4 connector, and 32-bit CompactPCI peripherals.

The backplane distributes PXI_CLK10, PXIe_CLK100, and PXIe_SYNC100 to every peripheral slot through independent buffers. PXI_CLK10 has less than 500 ps slot-to-slot skew, while PXIe_CLK100 is matched to less than 100 ps. The chassis also distributes eight shared PXI trigger lines for inter-module synchronization, sequencing, and event response.

External clocking is supported through the rear-panel 10 MHz REF IN BNC and through the PXI_CLK10_IN pin on the system timing slot. When a valid 10 MHz clock is present, the backplane automatically phase-locks PXI_CLK10, PXIe_CLK100, and PXIe_SYNC100 to that source. If clocks are present at both inputs, the system-timing-slot input takes priority. A copy of the backplane PXI_CLK10 is exported through the rear 10 MHz REF OUT connector.

The AC power system accepts 100–240 VAC at 50/60 Hz, with an operating range of 90–264 VAC at 47–63 Hz. The chassis provides up to 420 W of usable system power, with a 140 W maximum system-slot dissipation and 38.25 W maximum peripheral-slot dissipation. The fan controller adjusts speed based on rear intake-air temperature in AUTO mode, while the HIGH setting prioritizes cooling performance. At 0 to 30 °C in AUTO mode, the published sound-pressure level is 43.6 dBA.

The chassis is rated for 0 °C to 55 °C operation and 10% to 90% noncondensing relative humidity. It includes power-supply, temperature, and fan monitoring, and its modular power-supply shuttle is designed for field replacement. NI specifies that the chassis is not intended for life-safety-critical, hazardous-environment, or fail-safe applications.

 

Applications & System Role

The PXIe-1082 functions as the modular instrumentation enclosure, power supply, timing backbone, and PCIe switching fabric for a PXI Express test system. Its high-bandwidth backplane and dedicated x4 PCIe links make it suitable for systems combining high-throughput instruments, FPGA modules, data-acquisition modules, RF instruments, and controller hardware.

The four hybrid slots allow a mixed PXIe and hybrid-compatible PXI module population, which is useful when integrating existing PXI modules with newer PXI Express modules. The dedicated system timing slot supports a timing module for advanced synchronization and multi-chassis operation.

NI identifies the chassis as intended for high-performance test and measurement applications. Its extended 0 °C to 55 °C operating range, 38 W per-slot cooling, power/temperature/fan monitoring, and field-replaceable power-supply shuttle support demanding instrumentation installations.

 

FAQ

What is the NI PXIe-1082?
The NI PXIe-1082 is an eight-slot, 3U PXI Express chassis with a high-bandwidth backplane, AC power supply, and PXI timing and synchronization resources.

How many slots does the PXIe-1082 have?
It has eight physical slots: one system-controller slot, four hybrid peripheral slots, two PXI Express peripheral slots, and one PXI Express system timing slot.

Can it accept standard PXI modules?
Only hybrid-compatible standard PXI modules can be installed in the four hybrid slots, slots 5 through 8. The two PXI Express peripheral slots do not accept standard PXI modules.

What is the maximum system bandwidth?
The chassis supports up to 8 GB/s maximum system bandwidth and up to 2 GB/s single-direction dedicated bandwidth per PXI Express slot.

How much cooling does each slot provide?
Each slot provides 38 W of cooling capacity. The maximum peripheral-slot dissipation is 38.25 W.

What timing features does the chassis provide?
It provides PXI_CLK10, PXIe_CLK100, PXIe_SYNC100, eight PXI trigger lines, a PXI star trigger, and PXIe differential star triggers from the system timing slot.

Can the PXIe-1082 use an external 10 MHz clock?
Yes. It accepts a 10 MHz reference through the rear-panel REF IN connector or the PXI_CLK10_IN pin on the system timing slot, and it exports PXI_CLK10 through the rear REF OUT connector.

What are the operating limits?
The chassis operates from 0 °C to 55 °C and 10% to 90% noncondensing relative humidity, with an AC input of 100–240 VAC at 50/60 Hz.