Description
Core Product Summary
The NI SCXI-1300 is a general-purpose front-mount terminal block that provides convenient screw-terminal connections for 32-channel SCXI modules. It includes an onboard cold-junction temperature sensor for thermocouple measurements and attaches directly to the front of compatible modules via a 96-pin DIN connector.
The terminal block offers 72 screw terminals, including 32 pairs for differential analog inputs, chassis-ground connections, a safety-ground solder lug, and additional terminals for module OUTPUT, AOREF, and SCXIbus GUARD signals. It is designed for use with the SCXI-1100, SCXI-1102 series, SCXI-1104 series, SCXI-1181, and SCXI-1181K modules.
Key Technical Specifications
| Parameter | Specification |
|---|---|
| Product type | General-purpose front-mount terminal block |
| Compatible modules | SCXI-1100, SCXI-1102/B/C, SCXI-1104/C, SCXI-1181, SCXI-1181K |
| Number of screw terminals | 72 |
| Differential input pairs | 32 |
| Cold-junction sensor type | Integrated circuit (LM35CAZ) |
| Sensor accuracy | ±1.3 °C from 0 to 50 °C |
| Sensor repeatability | ±0.5 °C |
| Sensor output | 10 mV/°C (0 to 0.5 V from 0 to 50 °C) |
| Maximum field wire gauge | 26–16 AWG |
| Connector to module | 96-pin DIN |
| Operating temperature | 0 to 50 °C |
| Storage temperature | –20 to 70 °C |
| Relative humidity | 10 to 90 % noncondensing |
Detailed Product Overview
The SCXI-1300 serves as the primary field-wiring interface for selected 32-channel SCXI analog input modules. It mounts directly on the front of the module, converting the module’s high-density front connector into easily accessible screw terminals. This arrangement simplifies connection of thermocouples, voltage sources, and other low-level signals while providing the cold-junction compensation reference required for accurate thermocouple measurements.
Thirty-two pairs of screw terminals map directly to the differential input channels of the host module. Two additional terminals connect to the module’s functional chassis ground, and a solder lug provides a dedicated safety chassis ground connection. Extra terminals are available for the module’s analog output, analog output reference, and the SCXIbus GUARD signal, giving system designers flexible grounding and shielding options.
The onboard cold-junction sensor is an integrated-circuit device that produces a linear 10 mV/°C output. When the terminal block is used with a compatible module, the sensor voltage can be read either by multiplexing it with the analog input channels or by routing it to a dedicated DAQ device channel, depending on jumper configuration. Sensor accuracy is specified at ±1.3 °C over the 0 to 50 °C range, which includes both sensor tolerance and residual thermal gradients between the sensor and the screw terminals.
The terminal block is intended for indoor use and accepts field wiring from 26 to 16 AWG. It is fully shielded when properly installed on the host module, helping to maintain signal integrity in typical laboratory and industrial environments.

SCXI-1300

SCXI-1300
Dimensions, Mounting & Connection Notes
- Mounting method: Direct front attachment to compatible SCXI modules using the 96-pin DIN connector.
- Field connections: 72 screw terminals for signal, ground, and auxiliary connections.
- Wire size: 26–16 AWG.
- Safety ground: Dedicated solder lug for chassis safety grounding.
- Environment: Indoor use; operating temperature 0 to 50 °C.
System Compatibility & Integration Requirements
| Item / Dependency | Verified Compatibility or Requirement | Practical Note |
|---|---|---|
| Host modules | SCXI-1100, SCXI-1102/B/C, SCXI-1104/C, SCXI-1181, SCXI-1181K | Direct front-mount attachment via 96-pin DIN connector |
| Cold-junction compensation | Onboard IC sensor | Can be multiplexed with input channels or routed to a dedicated channel |
| Field wiring | 26–16 AWG | Strip insulation and secure under screw terminals |
| Grounding | Functional chassis ground terminals + safety ground solder lug | Use safety ground when required by local electrical codes |
| Mounting | Front of compatible SCXI module | Secured by the module’s front-panel fasteners |
FAQ
What is the NI SCXI-1300?
It is a general-purpose front-mount terminal block that provides screw-terminal access and an onboard cold-junction temperature sensor for selected 32-channel SCXI modules.
Which modules work with the SCXI-1300?
It is compatible with the SCXI-1100, SCXI-1102 series, SCXI-1104 series, SCXI-1181, and SCXI-1181K modules.
Does the SCXI-1300 include cold-junction compensation?
Yes. It contains an integrated-circuit temperature sensor with ±1.3 °C accuracy from 0 to 50 °C and a linear output of 10 mV/°C.
How many screw terminals does it have?
The terminal block provides 72 screw terminals, including 32 pairs for differential inputs plus ground and auxiliary connections.
What wire size is supported?
Field wiring from 26 to 16 AWG is accepted.
How is the terminal block mounted?
It attaches directly to the front of a compatible SCXI module via the 96-pin DIN connector and is secured by the module’s front-panel hardware.

WhatsApp: +86 16626708626
Email:
Phone: +86 16626708626