Sale!

Shinkawa VM-5G0-2 VM-5 Series Single Unit Instrument Rack

The Shinkawa VM-5G0-2 is a single-unit instrument rack from the VM-5 Series, designed to house and power individual monitor modules for rotating machinery protection. It integrates a built-in power supply and a preinstalled five-point relay module, providing a standalone solution for shaft vibration, displacement, or rotation monitoring without requiring a multi-slot chassis.

Categories: , , , , Brand:

Description

Core Product Summary

The Shinkawa VM-5G0-2 is a single-unit instrument rack from the VM-5 Series, designed to house and power individual monitor modules for rotating machinery protection. It integrates a built-in power supply and a preinstalled five-point relay module, providing a standalone solution for shaft vibration, displacement, or rotation monitoring without requiring a multi-slot chassis.

This unit delivers API 670 compliant alert and danger alarm contact outputs, featuring dry contact inputs for alarm reset and sequence operations. It supplies dedicated transducer power and supports direct integration with Shinkawa FK, VK, CV, or CA series transducers. The VM-5G0-2 operates within a 0 to 65 °C temperature range and includes plastic-sealed protective construction for industrial control environments.

 

Key Technical Specifications

Parameter Specification
Product Family Shinkawa VM-5 Series Monitor
Unit Type Single Unit Instrument Rack (Standalone)
Relay Configuration 5 points (DANGER1, ALERT1, DANGER2, ALERT2, OK)
Contact Rating 250 VAC, 0.2 A / 30 VDC, 2 A
Contact Life 100,000 operations minimum (rated load)
Transducer Power Output Input Code 1: 4 mA, 24 VDC
Input Code 2: -24 VDC, 20 mA
Input Code 3: 1CH: -24 VDC, 40 mA / 2CH: -24 VDC, 20 mA
Power Consumption 40 VA or less (VM-5G0 base configuration)
Operating Temperature 0 to 65 °C
Storage Temperature -30 to +85 °C
Relative Humidity 20 to 95% RH (non-condensing)
Insulation Resistance ≥ 100 MΩ at 500 VDC (Power supply to GND / GND to alarm contact)
Dielectric Strength 1,500 VAC for one minute (Power supply to GND)
Mass Rack: Max. 1.6 kg, Bezel: Max. 0.2 kg

 

Detailed Product Overview

The Shinkawa VM-5G0-2 functions as a standalone instrumentation hub within the VM-5 Series machinery protection system. Unlike multi-slot rack systems, this single-unit chassis integrates both the power supply and a preinstalled relay module, eliminating the need for separate power or relay expansion units. This design is optimized for localized monitoring points where a full 8-slot or 10-slot rack is unnecessary.

The unit is engineered to accommodate one VM-5 Series monitor module, such as dual vibration, thrust, or tachometer monitors. It directly supports the API 670 standard for alert and danger alarm contact outputs, ensuring compatibility with established turbine supervisory instrumentation (TSI) architectures. The built-in relay provides five distinct contact points (DANGER1, ALERT1, DANGER2, ALERT2, and OK), enabling direct sequence control integration without external relay modules.

Power distribution is managed internally, with configurable transducer power output options to match specific sensor requirements, including constant current or negative voltage supplies. The unit accepts dry contact inputs for external alarm reset and sequence operations, maintaining system state control from a central DCS or PLC. Its plastic-sealed protective construction and defined operating temperature range support deployment in standard industrial control rooms or protected machinery enclosures.

VM-5G0-2

VM-5G0-2

VM-5G0-2

VM-5G0-2

System Compatibility & Integration Requirements

Item / Dependency Verified Compatibility or Requirement Practical Note
Supported Monitor Units VM-5 Series Monitor units (e.g., VM-5K, VM-5T, VM-5S) Excludes VM-5F Temperature Monitor and Communication/Phase Marker units, which require multi-slot racks.
Power Supply Built-in (No external VM-5Z unit required) Base power consumption is 40 VA or less.
Relay Module Preinstalled 5-point SPDT relay Provides DANGER1, ALERT1, DANGER2, ALERT2, and OK outputs.
Transducer Compatibility Shinkawa FK, VK, CV, CA, RD series Power output must be matched to the sensor input code (1, 2, or 3).
Control System Interface Dry contact inputs Requires external normally open contacts for alarm reset and sequence operations.

 

Dimensions, Mounting & Connection Notes

  • Mounting Method: Panel mount (Single Unit Instrument Rack).
  • Mass: Maximum 1.6 kg for the rack, 0.2 kg for the bezel.
  • Connection Type: Terminal blocks for transducer power, alarm contacts, and dry contact inputs.
  • Environmental Limits: Operating temperature from 0 to 65 °C; storage from -30 to +85 °C. Relative humidity must remain between 20% and 95% RH, non-condensing.
  • Grounding and Insulation: Insulation resistance is maintained at 100 MΩ or more at 500 VDC between the power supply and ground, and between ground and alarm contacts.

 

FAQ

What is the Shinkawa VM-5G0-2?
The Shinkawa VM-5G0-2 is a single-unit instrument rack from the VM-5 Series, featuring a built-in power supply and a preinstalled 5-point relay module for standalone machinery protection monitoring.

Which monitor modules can be installed in the VM-5G0-2?
It accommodates a single VM-5 Series monitor unit, such as the VM-5K Dual Vibration Monitor or VM-5T Dual Thrust Monitor. It does not support the VM-5F Temperature Monitor or Communication/Phase Marker units.

What alarm outputs does the VM-5G0-2 provide?
It provides five SPDT relay contact points: DANGER1, ALERT1, DANGER2, ALERT2, and OK, which are compliant with API 670 standards for machinery protection.

What are the power requirements for the VM-5G0-2?
The unit has a built-in power supply with a maximum consumption of 40 VA (for the VM-5G0 base configuration). It also provides configurable DC power output to drive connected transducers based on the selected input code.

What is the operating temperature range for this instrument rack?
The specified operating temperature range is 0 to 65 °C, with a storage temperature range of -30 to +85 °C and relative humidity of 20 to 95% RH non-condensing.

+50